• Low profile plastic chamber, also suitable for patients with the same subcutaneous adipose tissue
    • Lightweight and affordable
    • Internal chamber in Titanium to prevent the needle from affecting the bottom and to ensure the absence of interaction with the infused substances
    • the large surface of the septum facilitates localization during skin palpation and reduces the risk of close punctures
    • transparent and radiopaque elastomeric connector: fast, safe and reversible connection of the catheter

    Caratteristiche

    Chamber
    Material Titanio + Polisulfone
    Shape and sizes: circular – diam. base. 27 mm
    Height: 10.6 mm
    Weight: 5.9 g
    Interior Volume : 0.36 ml

    Catetere con punta arrotondata, marcatura con punto ogni centimetro e numero ogni 5 centimetri.

    Septum
    Material Silicone
    Diameter available for punctures: 13 mm
    Maximum number of stings: 2000 with Huber needle of 22GA
    1000 with Huber  needle of 19GA
    Code
    Product
    description
    POWER TI
    Height
    mm
    Weight
    g
    Material
    ID
    mm
    OD
    mm
    Size
    Fr
    Lenght
    mm
    Introducer
    Dimension
    Fr
    Flow
    mL/min
    22GA
    Flow
    mL/min
    19GA
    201.22.12.21
    All-in-One
    10.6
    5.9
    silicone
    1.00
    2.10
    6,5
    780
    6.5
    7
    12
    201.22.12.31
    All-in-One
    10.6
    5.9
    silicone
    1.50
    2.70
    8
    780
    9
    10
    18
    201.22.12.41
    All-in-One
    10.6
    5.9
    PUR
    1.10
    1.70
    5
    780
    5
    8
    13
    201.22.12.20
    All-in-One
    10.6
    5.9
    PUR
    1.30
    2.00
    6,5
    780
    6.5
    7
    21
    201.22.12.33
    All-in-One
    10.6
    5.9
    PUR
    1.60
    2.70
    8
    780
    9
    8
    24
    201.22.10.21
    Standard
    10.6
    5.9
    silicone
    1.00
    2.10
    6,5
    780
    NA
    7
    12
    201.22.10.31
    Standard
    10.6
    5.9
    silicone
    1.50
    2.70
    8
    780
    NA
    10
    18
    201.22.10.41
    Standard
    10.6
    5.9
    PUR
    1.10
    1.70
    5
    780
    NA
    8
    13
    All-in-One Contents of Kit
    Chamber
    Catheter
    Connector
    Huber needle 22GA x 25.4 mm
    Luer connector
    Introducer-dilator peel-away
    Introducer needle from 18GA
    Guide to”J” of 0.89 mm
    Tunneler in stainless steel
    Standard Content of Kit
    Chamber
    Catheter
    Connector
    Huber needle of 22GA x 25.4 mm
    Luer Connector
    Venous introduction set
    Code Colour Code
    201.51.10.21 introduction set 6.5 FR
    10
    201.51.10.31 introduction set 9 FR
    10
    201.51.10.41 introduction set 5 FR
    10
    • Peel-away introducer-dilator
    • Introducer needle from 18 GA
    • Guide to “J” da 0.89 mm
    • Tunneling in stainless steel
  • PROCEDURAL PACKAGES

    For implants
    in extreme safety

    All our devices
    can be provided
    with a product kit
    for the plant to ensure
    an extremely safe procedure


  • PORT BRACHIALE


    Following the introduction of the ultrasound-guided venipuncture technique of the deep veins of the arm, it is now possible to combine the advantages of a fully implantable subcutaneous chamber (PORT) with the catheter introduction and placement technique typical of PICCs.
    The catheter is introduced and positioned as for the PICC implantation technique: ultrasound-guided venipuncture, micro introduction set, ECG control of the tip.
    The catheter is then connected to a small PORT chamber, inserted in a subcutaneous pocket of the middle third of the arm.

    Main indications:

    • chemotherapeutic treatments
    • anatomical difficulties with the implantation of traditional ports (prostheses, burns, tracheostomies, etc.)
    • aesthetic needs

    PICC-PORT main advantages:

    • absence of the exit-site
    • reduced risk of infections
    • less maintenance
    • reduced visibility of the chamber
    • less invasiveness and perceptibility for the patient

    Features

    Chamber
    Material: Titanio
    Shape and size: Wedge – base 21 mm x 16 mm
    Height: 8,6 mm
    Weight: 6 g
    Interior Volume: 0,2 ml
    Septum
    Material: Silicone

    Available diameter for punctures:
    9 mm
    Maximum number of stings: 1500 with Huber’s needle of 22GA

    Portata Max: 2 ml/s con Max pressione 300PSI, Viscosità Max 12cP


    Catheter with rounded tip, dot marking every centimeter and number every 5 centimeters

    Code
    Item
    description
    minMAX
    Height
    mm
    Weight
    g
    Material
    ID
    mm
    OD
    mm
    Size
    Fr
    Lenght
    mm
    Introductor
    Size
    Fr
    Flow
    ml/min
    22GA
    Flow
    ml/min
    19GA
    201.16.17.55
    All-in-one
    8.6
    6
    PUR
    1.10
    1.70
    5
    780
    5
    8
    1
     

    Kit Contents

    Chamber
    Catheter
    Connector
    Huber’s needle 22GA x 25.4 mm
    Luer’s connector
    Straight tip Nitinol guide
    Self-centered needle from 21 GA
    Dilator / Introducer
    Tunneler in stainless steel
  • 400 Ml

    600 Ml

    1000 Ml

    Central Venous Catheters

    • Complete range of Central Venous Catheters, single, double triple & quadruple lumen, meeting the needs of intensive care and emergency medication.
    • The Multi lumen catheters prevent mixing of incompatible drugs thus minimizing the complications.

     

    Kit Contents
    • Indwelling catheter
    • Introducer needle
    • Guide Wire
    • Guide wire pusher
    • Guide wire pusher tip cover
    • Guide wire pusher sleeve holder
    • Dilator
    • Luer lock syringe
    • Scalpel
    • Catheter holder
    • Catheter holder clamp
    • Injection Cap
    • Extension line Clamp

    Central Venous Catheter is available in various lumens and brands as mentioned in below table-

    Sr. No. Generic Name of Device Brand Name od device Product Code
    1 Central Venous Catheter Single Lumen Novocent 15001-15025, 15031-15049,15350
    Novocent pro 15500-15523,18000-18015, 15200-15223
    Novocent protekt 17000-17023, 18050-18064
    2 Central Venous Catheter Double Lumen Novocent 15050-15069,15071, 15085-15099,15361-15363
    Novocent pro 15530-15550, 18020-18034, 15225-15245
    Novocent protekt 17030-17050, 18070-18082
    3 Central Venous Catheter Triple Lumen Novocent 15100-15117, 15150-15163
    Novocent pro 15560-15579, 18040-18048, 15246-15265,15280-15281
    Novocent protekt 17055-17073, 18085-18093
    4 Central Venous Catheter Quadruple Lumen Novocent 17271-17272
    Novocent pro 15266-15277, 15601-15612
    Novocent protekt 17260-17270, 18100-18102, 18110-18112

    Common Features

    • Central Venous Catheter made of specially formulated and bio-compatible Polyurethane material provides strength during insertion.
    • Polyurethane softens at body temperature, conforms to the body tissues and reduces the risk of vascular trauma.
    • Specially designed soft tip for smooth & easy insertion of catheter.
    • Soft Flexible J-Tip Guidewire prevents the vessel perforation and also provides good torque to ensure firm insertion.
    • Radio-opaque material of catheter with clear, definite marking facilitates correct placement of catheter tip.
    • Novocent available with J-tip guidewire made up of stainless steel.
    • Novocent Pro available with kink resistant flexible guidewire made up of Nitinol with J-tip and lipid resistant three way stop cock.
    • Novent Protekt available with kink resistant flexible guidewire made up of Nitinol with J-tip, lipid resistant three way stop cock and clave connector.

    Options Available

    • Y-Introducer Needle
    • Kink resistant guidewire
    • Clave Connector
    • Three way stop cock
    • Guiding syringes

    • Double room with different light tubes
    • Differentiated diameters and heights for safe subcutaneous recognition of each chamber
    • Suitable for infusion of incompatible substances (e.g. risk of crystallization in case of mutual contact)
    • Large self-sealing silicone septum
    • Internal chamber in titanium to prevent the needle from affecting the bottom and to ensure the absence of interaction with the infused substances

    Features

    Chamber
    Material Titanio + Polisulfone
    Shape and size: base 50 mm x 34 mm
    Height: 13.9 mm – 15.4 mm
    Weight: 19.7 g
    Internal volume: 0.53 ml – 0.52 ml

    Catheter with rounded tip, dot marking every centimeter and number every 5 centimeters.

    Septum
    Material Silicone
    Available diameter for punctures *: 12.7 mm
    Maximum number of stings: 2000 with Huber needle from 22GA
    1000 with Huber needle from 19GA
    * For each septum

    Standard

    Code
    Catheter
    ø int
    mm
    ø ext
    mm
    ø ext
    mm

    lenght
    FR

    Volume
    ml

    201.21.10.91
    Silicone to double lumen
    small 1.0
    big 1.3
    3.4
    10.5
    600
    small 0.61
    big 1.3
    Kit Contents
    Chamber
    Catheter
    Connector
    Huber’s needle  22GA x 25.4 mm
    Luer Connector
    Code
    Item
    description
    POWER TI
    Height
    mm
    Weight
    g
    Material
    ID
    mm
    OD
    mm
    Size
    Fr
    Lenght
    mm
    Introducer
    Dimension
    Fr
    Flow
    mL/min
    22GA
    Flow
    mL/min
    19GA
    201.15.12.21
    All-in-One
    10.2
    8.5
    silicone
    1.00
    2.10
    6,5
    780
    6.5
    7
    12
    201.15.12.31
    All-in-One
    10.2
    8.5
    silicone
    1.50
    2.70
    8
    780
    9
    10
    18
    201.15.12.41
    All-in-One
    10.2
    8.5
    PUR
    1.10
    1.70
    5
    780
    5
    8
    13
    201.15.10.21
    Standard
    10.2
    8.5
    silicone
    1.00
    2.10
    6,5
    780
    NA
    7
    12
    201.15.10.31
    Standard
    10.2
    8.5
    silicone
    1.50
    2.70
    8
    780
    NA
    10
    18
    201.15.10.41
    Standard
    10.2
    8.5
    PUR
    1.10
    1.70
    6,5
    780
    NA
    8
    13
    All-in-One  Kit Contens
    Chamber
    Catheter
    Connector
    Huber’s needle 22GA x 25.4 mm
    Luer’s connector
    Introductor-dilatator peel-away
    Introducer needle of 18GA
    Guide of “J” from 0.89 mm
    Tunneler in stainless steel

    Venous introduction set

    Code Colour Code
    201.51.10.91 Introduction set 11 FR
    10
    • Introducer-dilator peel-away
    • Introducer needle from 18 GA
    • Guide to “J” of 0.89 mm
    • Tunneling in stainless steel
  • Spinal System

    • Small size
    • “Wedge” shape to facilitate insertion into the subcutaneous pocket
    • The large surface of the septum facilitates localization during skin palpation and reduces the risk of close punctures
    • Excellent stability of the needle due to the high compression of the septum allowed by the Titanium chamber
    • Titanium guarantees maximum reliability and mechanical stability

    Caratteristiche

    Chamber
    Material Titanio
    Shape and Dimension: Wedge – base 27.5 mm x 19 mm
    Height: 10.5 mm
    Weight: 10.5 g
    Interior Volume: 0.53 ml
    Septum
    Material Silicone
    Diameter available for stings: 12.8 mm
    Maximum number of stings: 2000 with Huber’s needle of 22GA
    1000 with huber’s needle of 129GA

    A 50 micron filter inside the chamber prevents clogging of the system due to small particles

    Code
    Catheter
     Chamber
    Ø int
    mm
    Ø ext
    mm
    Ø ext
    GA
    Lenght
    mm
    Volume
    ml
    201.17.10.61
    ETI
    Open toe PU
    0.5
    1.0
    19
    900
    0.18
    201.17.10.71
    ETI
    Closed toe PU
    and side holes
    0.5
    1.0
    19
    900
    0.18
    Kit Contents
    Chamber
    Catheter
    Connector
    Huber’s needle  22GA x 25.4 mm
    Needle to tip Tronca
    Tuohy’s needle 17GA x 80 mm
    Tunneller in stainless stell

    Arterial System

    • Small size
    • “Wedge” shape to falicitate insertion into the subcutaneous pocket
    • The large surface of the septum facilitates localization during skin palpation and reduces the risk of close punctures
    • Excellent stability of the needle due to the high compression of the septum allowed by the Titanium chamber
    • Titanium guarantess maximum realibility and nechanical stability

    Features

    Chamber
    Material Titanio
    Shape and size: Wedge – base 27.5 mm x 19 mm
    Height: 10.5 mm
    Weight: 10.5 g
    Interior volume: 0.53 ml
    Septum
    Material Silicone
    Available dimater for punctures
    12.8 mm
    Maximum number of stings: 2000 with Huber’s needle of 22GA
    1000 with Huber’s needle of 129GA
    Code
    Chamber
    Catheter
    mm
    Ø int
    mm
    Ø ext
    mm
    Ø ext
    GA
    Lenght
    mm
    Volume
    ml
    201.17.10.51
    ETI
    Silicone
    1.0
    2.1
    6.5
    780
    0.61
    Kit contents
    Chamber
    Catheter
    Connector
    Huber’s needle  22GA x 25.4 mm
    Luer’s connector
    • Small size
    • “Wedge” shape to facilitate insertion into the subcutaneous pocket
    • Large septal surface facilitates localization during skin palpation and reduces the risk of close punctures
    • Excellent needle stability due to the high septal compression enabled by the Titanium chamber
    • Titanium ensures maximum mechanical reliability and stability

    Details

    Chamber
    Material Titanium
    Shape and size: Wedge – base 27.5 mm x 19 mm
    Height: 10.5 mm
    Weight: 10.5 g
    Internal Volume: 0.53 ml
    Septum
    Material Silicone
    Diameter available for punctures: 12.8 mm
    Maximum number of punctures: 2000 with 22GA Huber’s needle
    1000 with 129GA Huber’s needle

    Catheter with rounded tip, marking with dot every centimeter and number every 5 centimeters.

    Code
    Description
    product
    ETI
    Height
    mm
    Weight
    g
    Material
    ID
    mm
    OD
    mm
    Size
    Fr
    Lenght
    mm
    Introducer
    Size
    Fr
    Scope
    mL/min
    22GA
    Scope
    mL/min
    19GA
    201.17.12.21
    All-in-One
    10.5
    10.5
    silicone
    1.00
    2.10
    6,5
    780
    6.5
    7
    13
    201.17.12.31
    All-in-One
    10.5
    10.5
    silicone
    1.50
    2.70
    8
    780
    9
    10
    18
    201.17.12.41
    All-in-One
    10.5
    10.5
    PUR
    1.10
    1.70
    5
    780
    5
    8
    13
    201.17.10.21
    Standard
    10.5
    10.5
    silicone
    1.00
    2.10
    6,5
    780
    NA
    7
    13
    201.17.10.31
    Standard
    10.5
    10.5
    silicone
    1.50
    2.70
    8
    780
    NA
    10
    13
    201.17.10.41
    Standard
    10.5
    10.5
    PUR
    1.10
    1.70
    5
    780
    NA
    8
    13
    All-in-One Kit Contents
    Chamber
    Catheter
    Connector
    Huber 22GA x 25.4 mm needle
    Luer Connector
    Peel-away introducer-dilator
    18GA introducer needle
    0.89 mm “J” guide
    Stainless steel tunneler
    Standard Kit Contents
    Chamber
    Catheter
    Connector
    Huber 22GA x 25.4 mm needle
    Luer Connector

    Venous Introduction Set

    Code Colour Code
    201.51.10.21 Introduction set 6.5 FR
    10
    201.51.10.31 Introduction set 9 FR
    10
    201.51.10.41 Introduction set 5 FR
    10
    • Peel-away introducer-dilator
    • 18 GA introducer needle
    • 0.89 mm “J” guide
    • Stainless steel tunneling
    • Chamber designed to be used at high flows for injections of contrast media
    • The large surface of the septum facilitates localization during skin palpation and reduces the risk of close punctures
    • Transparent and radiopaque elastometric connector: fast, safe and reversible connection of the catheter
    • Low profile plastic chamber, also suitable for patients with thick subcutaneous adipose tissue
    • Internal chamber in Titanium to prevent the needle from affecting the bottom and to ensure the absence of interaction with the infused substances

    Caratteristiche

    Chamber
    Material Titanio + Polisulfone
    Shape and size: Circular – diam. base. 27 mm
    Height: 10.6 mm
    Weight: 5.9 g
    Internal volume: 0.36 ml
    Septum
    Material Silicone
    Diameter available for punctures: 13 mm
    Maximum number of stings: 2000 with 22GA Huber needle
    1000 with 19GA Huber needle

    Max Flow Rate: 5ml / s with Max pressure 300PSI, Max Viscosity 12cP

    Code
    Item
    description
    POWER PLP
    Height
    mm
    Weight
    g
    Material
    ID
    mm
    OD
    mm
    Dimension
    Fr
    Lenght
    mm
    Dimensions
    introducer
    Fr
    Flow
    mL/min
    22GA
    Flow
    mL/min
    19GA
    201.24.12.30
    All-in-One
    10.6
    5.9
    silicone
    1.30
    2.70
    8
    600
    9
    7
    21
    201.24.12.20
    All-in-One
    10.6
    5.9
    PUR
    1.35
    2.00
    6
    600
    6.5
    8
    22
    201.24.12.33
    All-in-One
    10.6
    5.9
    PUR
    1.60
    2.70
    8
    600
    9
    8
    24
    201.24.10.30
    Standard
    10.6
    5.9
    silicone
    1.30
    2.70
    8
    600
    NA
    7
    21
    201.24.10.20
    Standard
    10.6
    5.9
    PUR
    1.35
    2.00
    6
    600
    NA
    8
    22
    201.24.10.33
    Standard
    10.6
    5.9
    PUR
    1.60
    2.70
    8
    600
    NA
    8
    24
    All-in-One Kit contents
    Chamber
    Catheter
    Connector
    Huber’s needle 22GA x 25.4 mm
    Luer’s connector
    Introducer-dilator peel-away
    Introducer needle da 18GA
    Guide at “J” da 0.89 mm
    Tunneler in stainless steel
    Standard Standard Kit Contents
    Chamber
    Catheter
    Connector
    Huber’s needle 22GA x 25.4 mm
    Luer’s connector

    Venous introduction set

    Code Colour Code
    201.51.10.21 Introduction set 6.5 FR
    10
    201.51.10.31 Introduction set 9 FR
    10
    • Introducer-dilator peel-away
    • Introducer needle from 18 GA
    • Guide at “J” da 0.89 mm
    • Tunneler in stainless steel
    • Chamber designed to be used at high flows for injections of contrast media
    • “Low-Profile” chamber to facilitate insertion into the subcutaneous pocket and increase patient comfort
    • Emerging domed septum for immediate and safe subcutaneous retrieval
    • The large surface area of ​​the septum reduces the risk of close punctures
    • Transparent and radiopaque elastomeric connector: fast, safe and reversible connection of the catheter

    Caratteristiche

    Chamber
    Material Titanio
    Shape and size: Circular – diam. base. 25 mm
    Height: 10 mm
    Wheight: 10.3 g
    Internal Volume: 0.6 ml

    Catheter with rounded tip, dot marking every centimeter and number every 5 centimeters.

    Septum
    Material Silicone
    Diameter available for punctures: 12.8 mm
    Maximum number of stings: 2000 with Huber’s needle from 22GA
    1000 with Huber’s needle from 19GA

    Max Flow Rate: 5ml / s with Max pressure 300PSI, Max Viscosity 12cP

    Code
    Item
    description
    POWER LPT
    Height
    mm
    Weight
    g
    Material
    ID
    mm
    OD
    mm
    Size
    Fr
    Lenght
    mm
    Introductor
    Size
    Fr
    Flow
    mL/min
    22GA
    Flow
    mL/min
    19GA
    201.26.12.30
    All-in-One
    10
    10.3
    silicone
    1.30
    2.70
    8
    600
    9
    7
    21
    201.26.12.20
    All-in-One
    10
    10.3
    PUR
    1.35
    2.00
    6
    600
    6.5
    8
    22
    201.26.12.33
    All-in-One
    10
    10.3
    PUR
    1.60
    2.70
    8
    600
    9
    8
    24
    201.26.10.30
    Standard
    10
    10.3
    silicone
    1.30
    2.70
    8
    600
    NA
    7
    21
    201.26.10.20
    Standard
    10
    10.3
    PUR
    1.35
    2.00
    6
    600
    NA
    8
    22
    201.26.10.33
    Standard
    10
    10.3
    PUR
    1.60
    2.70
    8
    600
    NA
    8
    24
    All-in-One Kit content
    Chamber Needle truncated point
    Catheter Syringe 10 ml
    Coonector
    Huber’s needle 22GA x 25.4 mm Huber’s needle 90° – 20GA x1″ with extension (in dedicated packaging)
    Luer’s connector
    Introducer-dilator peel-away
    ntroducer needle from 18GA
    Guide at “J” da 0.89 mm
    Tunneller in stainless steel
    Standard Kit of content
    Chamber
    Catheter
    Connector
    Huber’s needle 22GA x 25.4 mm
    Luer’s connector

    Venous introduction set

    Code Colour Code
    201.51.10.21 Introduction Set 6.5 FR
    10
    201.51.10.31 Introduction Set 9 FR
    10
    • Introducer-dilator peel-away
    • Introducer needle from 18 GA
    • Guide at “J” da 0.89 mm
    • Tunneling in stainless steel
    • High profile
    • Chamber designed to be used at high flows for injections of contrast media
    • The large surface of the septum facilitates localization during skin palpation and reduces the risk of close punctures
    • Transparent and radiopaque elastomeric connector: fast, safe and reversible connection of the catheter

    Caratteristiche

    Chamber
    Material Titanio
    Shape and size: Wedge- diam. base. 22 mm x 26 mm
    Height: 13 mm
    Wheight: 12 g
    Internal Volume: 0.82 ml

    Catheter with rounded tip, dot marking every centimeter and number every 5 centimeters.

    Septum
    Material Silicone
    Diameter available for punctures: 13 mm
    Maximum number of stings: 2000 with Huber’s needle from 22GA
    1000 with Huber’s needle from 19GA

    Max Flow Rate: 5ml / s with Max pressure 300PSI, Max Viscosity 12cP

    Code
    Item
    description
    POWER TI
    Height
    mm
    Weight
    g
    Material
    ID
    mm
    OD
    mm
    Size
    Fr
    Lenght
    mm
    Introducer
    Dimension
    Fr
    Flow
    mL/min
    22GA
    Flow
    mL/min
    19GA
    201.23.12.30
    All-in-One
    13
    12
    silicone
    1.30
    2.70
    8
    600
    9
    7
    21
    201.23.12.20
    All-in-One
    13
    12
    PUR
    1.35
    2.00
    6
    600
    6.5
    8
    22
    201.23.12.33
    All-in-One
    13
    12
    PUR
    1.60
    2.70
    8
    600
    9
    8
    24
    201.23.10.30
    Standard
    13
    12
    silicone
    1.30
    2.70
    8
    600
    NA
    7
    21
    201.23.10.20
    Standard
    13
    12
    PUR
    1.35
    2.00
    6
    600
    NA
    8
    22
    201.23.10.33
    Standard
    13
    12
    PUR
    1.60
    2.70
    8
    600
    NA
    8
    24
    All-in-One Kit Contents
    Chamber
    Catheter
    Connector
    Huber’s needle 22GA x 25.4 mm
    Luer Connector
    Introducer-dilator peel-away
    Introducer needle from da 18GA
    Guide at “J” from 0.89 mm
    Tunneler in stainless steel
    Standard Kit Contents
    Chamber
    Catheter
    Connector
    Huber’s needle 22GA x 25.4 mm
    Luer

    Venous introduction set

    Code Colour Code
    201.51.10.21 Introduction Set 6.5 FR
    10
    201.51.10.31 Introduction Set 9 FR
    10
    • Introducer-dilator peel-away
    • Introductor needle da 18 GA
    • Guide at “J” from 0.89 mm
    • Tunneling in stainless steel